Oscilloscope Familiarization Labs
At this point, you should see a sine wave similar to
. But if you look
closely you should also see some glitches (narrow pulses) near the peaks of this
sine wave. And the amplitude of these glitches may appear to vary (bouncing up
and down). The amplitude of these glitches is actually very stable. The problem is
that the scope has reduced its sample rate (note the sample rate shown under the
Keysight logo on the scope’s display) and the scope is now intermittently
capturing the narrow glitches. The scope is under-sampling the narrow glitches.
Sometimes the scope captures a single point on the peak of a glitch. Sometimes it
captures a point on a transition of the glitch. And sometimes it captures nothing at
all on the glitch (the glitch width is narrower than the sample interval). This scope
has a special acquisition mode called “Peak Detect” that will resolve this problem.
Let’s turn it on.
8
Press the
[Acquire]
front panel key (below the Cursors knob).
9
Press the
Acq Mode
softkey and select
Peak Detect
.
Figure 13
The scope’s automatically-reduced sample rate under-samples the repetitive
glitch.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...