Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
4
4000 X-Series Oscilloscopes Advanced Training Guide
121
Using the default trigger value of arming on right-channel value
≤
-10 and then
trigger on right-channel value
≥
+10, your scope should now be triggering on “R:
+20” with this byte shown at center-screen similar to
. Let’s now capture
a longer stream of I
2
S traffic and then manually search through the captured and
decoded record.
25
Press the
[Serial]
front panel key.
26
Tap the
Lister
softkey; then use the
Window
menu to select
Half-Screen
.
27
Set the scope’s timebase to
10.00 ms/div
.
28
Press
[Run/Stop]
to stop repetitive acquisitions.
29
Tap the
Scroll Lister
softkey; then turn the Entry knob to manually scroll through
the lister table.
As you scroll through the data, note that waveforms “track”. This means that the
frame that the arrow points to in the lister table corresponds to the waveforms that
are positioned at center-screen. If you want to zoom in on a particular I
2
S frame of
data, either push the Entry knob, or tap the
Zoom to Selection
softkey. Let’s now
perform an automatic search to find every occurrence of Right-channel > +10. We
will then automatically navigate to each of these occurrences.
30
Push the horizontal position/delay knob to re-position the trigger point back to
center-screen.
Figure 81
Triggering on I
2
S right-channel = +20.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...