Performance Verification Tests
4
Keysight 3458A Calibration Manual
65
Analog AC voltage performance tests
Required equipment
The following list of equipment is required to test the analog AC performance of
the 3458A.
– Stable AC voltage source (Fluke 5700A or equivalent).
– Transfer Standard DMM (3458A Opt. 002 within 90 days of Cal.)
– Shielded test leads terminated with dual banana plugs (such as
Keysight 11000-60001).
Preliminary steps
1
Make certain that the AC source is properly warmed up.
2
The 3458A requires a 4 hour warm up period. If this has not been done, turn
the instrument ON and allow it to warm up.
3
Execute the ACAL AC function on both the transfer standard DMM and the
UUT. This auto calibration will take approximately 1 minute to complete.
4
If the instrument Self Test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF
TEST PASSED”.
5
Configure the transfer standard DMM as follows:
– ACV
– SETACV SYNC
– ACBAND 10,2E6
– RANGE10
– RES .001
– TRIG SGL
– LFILTER ON
6
Configure the DMM under test as follows:
– ACV
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...