Adjustment Procedures
3
Keysight 3458A Calibration Manual
43
Figure 3-2
DC gain adjustment connections
Resistance and DC current adjustment
This adjustment calculates gain corrections for the resistance and DC current
ranges. The DC Gain Adjustment must be performed prior to this adjustment
because this adjustment relies on the values calculated by the DC Gain
Adjustment.