Using N8480 Series Power Sensors
6
Keysight N1913/1914A User’s Guide
187
Sensor Calibration Tables
This section describes how to use sensor calibration tables. Sensor calibration
tables store the measurement calibration factors, for a power sensor model or for
a specific power sensor, in the power meter. They are used to correct
measurement results.
Use sensor calibration tables when you want to make power measurements over a
range of frequencies using one or more power sensors.
The N1913/1914A EPM Series power meters are capable of storing 20 sensor
calibration tables, each containing up to 80 frequency points. The power meter is
supplied with a set of nine predefined sensor calibration tables plus a “100%”
default table. The data in these tables is based on statistical averages for a range
of Keysight Technologies power sensors. Your own sensor will most likely differ
from the typical to some degree. If you require best accuracy, create a custom
table for each sensor you use as shown in
“Editing/Generating Sensor Calibration
To use calibration factor tables,
1
Select the sensor calibration table to be applied to a channel.
2
Zero and calibrate the power meter. The reference calibration factor used
during the calibration is automatically set by the power meter from the sensor
calibration table.
3
Specify the frequency of the signal you want to measure. The calibration factor
is automatically set by the power meter from the sensor calibration table.
4
Make the measurement.
Содержание 1914A
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters User s Guide...
Страница 24: ...24 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 54: ...1 Introduction 54 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 206: ...7 Using U2000 Series USB Power Sensors 206 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 240: ...9 Maintenance 240 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 242: ...10 Characteristics and Specifications 242 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...