5
Using 8480 Series Power Sensors
160
Keysight N1913/1914A User’s Guide
Figure 5-3
Reference Calibration Factor pop-up window
5
Press
to complete the entry.
Now, zero and calibrate the power meter/sensor combination as follows:
6
Press
and the channel
softkey. The
Zeroing
pop-up is displayed.
7
Connect the power sensor to the POWER REF output.
8
Press
and the channel
softkey to start the calibration routine. The
Calibrating
pop-up is then displayed.
Now, set the sensor calibration factor for the frequency of the signal you want to
measure.
9
Check the current reference calibration factor setting by pressing
,
. The value is displayed on the
Cal Fac
field.
%
Zero
Cal
NOTE
You can reduce the steps required to carry out the zero and calibration
procedure as follows:
– Connect the power sensor to the POWER REF output.
– Press
and
. (For dual channel meters, press
,
or
as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
Offset
NOTE
Does this setting match the value for the sensor? (The power sensor reference
calibration factor can normally be found above the calibration factors table on
the power sensor body.)
Содержание 1914A
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters User s Guide...
Страница 24: ...24 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 54: ...1 Introduction 54 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 206: ...7 Using U2000 Series USB Power Sensors 206 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 240: ...9 Maintenance 240 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 242: ...10 Characteristics and Specifications 242 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...