3
Operating Instructions
28
Keysight 11683A Operating and Service Manual
Operation of the Controls
Operation of the controls of the 11683A is explained in
.
Figure 3-1
Front and rear panel features (illustrations for indication only)
1 LINE switch
. Controls primary power. Illuminated when instrument is ON.
2 RANGE switch
. Equivalent to compatible Power Meter's Range Switch; produces a full scale
Power Meter reading when 11683A and Power Meter Range switches are set to same scale.
3 FUNCTION switch
. When the switch is set to CALIBRATE an output dependent on the RANGE
switch setting is coupled to the Power Meter. In STANDBY mode the output is grounded.
4 POWER METER connector
. Connects the output to, and control signals from, compatible
Power Meter via Power Sensor Cable.
5 POLARITY switch
. An upscale reading is obtained on the Power Meter when the switch is set to
NORMAL. The REVERSE setting produces a down
scale reading.
6 D.C. REFERENCE OUTPUT connector
. DC reference voltage output from RANGE
Switch. Load resistance must be
≥
100 M
Ω
for proper operation of the 11683A.
7 Power Module Assembly
.
8 Receptacle
. Couples transformer primary to line voltage via power cable.
9 Line Voltage Selection Card
. Matches transformer primary to line voltage. Refer to
10 Fuse
. A 1/8 A fuse is used at 100/120 Vac; 1/16 A fuse at 230/240 Vac.
11 Fuse Pull Handle
. Mechanical interlock; fuse must be removed before extraction of Line Voltage
Selection Card.
12 Window
. Safety interlock; fuse cannot be removed while power cable is coupled to Power
Module Receptacle.
Содержание 11683A
Страница 1: ...Keysight 11683A Range Calibrator Operating and Service Manual...
Страница 12: ...12 Keysight 11683A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 14: ...14 Keysight 11683A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...2 Installation 26 Keysight 11683A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 48: ...6 Service 48 Keysight 11683A Operating and Service Manual Figure 6 3 11683A assembly and component locations...
Страница 51: ...Service 6 Keysight 11683A Operating and Service Manual 51 Figure 6 11 11683A FG top view...