Keysight 1158 Series Скачать руководство пользователя страница 26

26

Keysight 849x and 1158x Operating and Service Manual

8493A/B

The 8493A/B has a male SMA jack on one end and a female SMA on the other. These connectors 

mate with the opposite sex SMA connectors

Figure 5

SMA connector diagram

Содержание 1158 Series

Страница 1: ...Keysight 8490D G 8491A B 8493A B C Coaxial Fixed Attenuators Keysight 11581A 11582A 11583C Attenuator Sets Operating and Service Manual...

Страница 2: ...hese rights customarily provided to the public to use modify reproduce release perform dis play or disclose commercial computer soft ware or commercial computer software documentation No additional go...

Страница 3: ...or electronic product in domestic household waste Product category With reference to the equipment types in the WEEE directive Annex 1 this instrument is classified as a Monitoring and Control Instru...

Страница 4: ...4 Keysight 849x and 1158x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 5: ...ral Information 11 Attenuator overview 11 Features 11 Optional calibration data 12 Attenuator sets 12 Specifications 13 Environmental Specifications 19 Installation 20 Initial inspection 20 Returning...

Страница 6: ...6 Keysight 849x and 1158x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 7: ...ures Figure 1 Equipment setup using oscillator and SWR meter 21 Figure 2 Equipment setup using network analyzer 23 Figure 3 2 4 mm 1 85mm connector diagram 24 Figure 4 Type N connector diagram 25 Figu...

Страница 8: ...8 Keysight 849x and 1158x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 9: ...ttenuator 14 Table 5 Specifications for 8491A and 8493A coaxial fixed attenuators 15 Table 6 Specifications for 8491B and 8493B coaxial fixed attenuators 16 Table 7 Specifications for 8493C coaxial fi...

Страница 10: ...10 Keysight 849x and 1158x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 11: ...ations Accuracy and low SWRs make the attenuators well suited for extending the range of sensitive power meters for higher power measurements and applications such as calibration standards and RF subs...

Страница 12: ...hnologies Service Centers Attenuator sets The 11581A 11582A and 11583C are boxed sets of four coaxial fixed attenuators of 3 6 10 and 20 dB as shown in Table 2 Each attenuator set comes with a printed...

Страница 13: ...3 indicates 3 dB attenuation Option 010 indicates 10 dB attenuation and so on DC to 50 GHz DC to 26 5 GHz 26 5 to 50 GHz 003 2 5 3 9 4 8 006 5 4 6 9 7 8 010 9 4 10 9 11 3 020 19 2 21 3 21 7 030 29 2 3...

Страница 14: ...1 5 31 7 040 38 0 42 5 42 5 42 5 Maximum SWR Option a a Option numbers indicate the attenuation values For example Option 003 indicates 3 dB attenuation Option 010 indicates 10 dB attenuation and so o...

Страница 15: ...003 0 3 006 0 3 010 0 5 020 0 5 030 1 0 040 b b Options 040 050 and 060 are not available for 8493A 1 5 050 b 1 5 060 b 2 0 SWR Option a DC to 8 GHz 8 to 12 4 GHz 003 1 25 1 35 006 1 2 1 3 010 1 2 1 3...

Страница 16: ...on 003 indicates 3 dB attenuation Option 010 indicates 10 dB attenuation and so on DC to 26 5 GHz 26 5 to 40 GHz 40 to 50 GHz 003 1 25 1 3 1 5 006 1 2 1 3 1 5 010 1 2 1 3 1 5 020 1 2 1 3 1 5 030 1 2 1...

Страница 17: ...a DC to 18 GHz 18 to 26 5 GHz 003 0 5 1 0 006 0 6 0 6 010 0 3 0 5 020 0 5 0 6 030 0 7 1 0 040 1 0 1 3 SWR Option a a Option numbers indicate the attenuation values For example Option 003 indicates 3 d...

Страница 18: ...ements S11 S22 Attenuation DC to 4 GHz 4 to 12 GHz 12 to 18 GHz 3 dB 0 01 dB 0 06 dB 0 11 dB 6 dB 0 01 dB 0 07 dB 0 11 dB 10 dB 0 01 dB 0 07 dB 0 12 dB 20 dB 0 01 dB 0 08 dB 0 13 dB Female Male Model...

Страница 19: ...fully comply with Keysight Technologies product operating environmental specifications as shown Table 10 Table 10 849x coaxial fixed attenuator and 1158x attenuator sets environmental specifications T...

Страница 20: ...ent Keep the shipping materials for the carrier s inspection 3 If you are returning the instrument under warranty or for service repackaging the instrument requires original shipping containers and ma...

Страница 21: ...rocedure 1 Connect the equipment as shown in the Figure 1 except remove the attenuator and connect the adapters directly together Figure 1 Equipment setup using oscillator and SWR meter 1 Set the test...

Страница 22: ...dB Meter indication dB Minimum Actual Maximum 0 system cal 2 or 10 a a Set SWR meter range to 10 dB and power level to 0 0 dB for 3 dB 6 dB and 10 dB attenuators only Set to 0 5 or 0 0 a 3 10 1 35 1 6...

Страница 23: ...ports of the network analyzer This depends on the type of the attenuator being checked The equipment setup is as illustrated in Figure 2 1 Calibrate the network analyzer with appropriate settings and...

Страница 24: ...d by other devices connected to them The attenuators are bidirectional that is the signal may be inserted from either end 8490D G The 8490D 2 4 mm connectors mate with other 2 4 mm connectors of the o...

Страница 25: ...rvice Manual 25 8491A B The 8491A B type N connectors mate with all type N connectors whose dimensions conform to IEE STD 287 Figure 4 Type N connector diagram NOTE Do not mate with 0 071 inch diamete...

Страница 26: ...49x and 1158x Operating and Service Manual 8493A B The 8493A B has a male SMA jack on one end and a female SMA on the other These connectors mate with the opposite sex SMA connectors Figure 5 SMA conn...

Страница 27: ...3C has a male 3 5 mm connector on one end and a female 3 5 mm connector on the other side These connectors mate with the opposite sex 3 5 mm or SMA connectors Figure 6 3 5 mm connector diagram NOTE Co...

Страница 28: ...ormance Test The attenuators can be tested to the accuracy of the specifications with a network analyzer or equivalent equipment of suitable accuracy If a network analyzer is available test the instru...

Страница 29: ...ir since the cards must be mounted in cartridges to test and testing costs more than a replacement attenuator Maintenance The connectors particularly the connector faces must be kept clean This is esp...

Страница 30: ...30 Keysight 849x and 1158x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 31: ...is subject to change without notice Always refer to the Keysight website for the latest revision Keysight Technologies 2014 2018 Edition 6 November 27 2018 Printed in Malaysia 08491 90077 08491 90077...