Performance Verification for 1147B Probe
17
Performance Verification for 1147B Probe
Use the following procedure to test the warranted Accuracy
specifications (as listed on
page 21
) for the 1147B probe. The
recommended test interval to warrant performance of this probe is
once a year. However, you can also test the probe’s performance
as and when required using the recommended test equipment and
by following the procedure documented in this chapter.
Table 1
Required Test Equipment
Description
Minimum Requirements
Recommended Test Equipment
Oscilloscope
Bandwidth
≥
250 MHz
Amplitude accuracy:
≤
0.4%
Infiniium oscilloscope with 1 M
Ω
input
Two Digital Multimeters
-
Keysight U1251B Handheld
Digital Multimeter or equivalent
DVM/DMM
AC Current Generator
AC 10 A
rms
, 50 Hz, sine wave
Amplitude accuracy:
≤
0.3%
Keysight AC6800 or Wavetek
9100
Probe Adapter
-
Keysight N1022B Probe Adapter
Power Supply
-
Keysight 1143A Probe Offset
Control and Power Module
Test Cables (with banana jacks) -
-
BNC Cable
-
-
BNC (F) to Banana Adapter
-
-
NMD 3.5mm (M) to 3.5mm (F)
adapter
-
-
3.5mm (M) to BNC (F) adapter
-
-