To test trigger sensitivity
This test verifies the trigger sensitivity. In this test, you apply a sine wave to the
oscilloscope at the bandwidth limits. You then decrease the amplitude of the
signal to the specified levels, and check to see if the oscilloscope is still triggered.
Table 10
Trigger Sensitivity Specification
Trigger sensitivity Ch 1, 2, 3, 4 (DC
coupling):
≥
5 mV/div: 1 div from DC to 10 MHz, 1.5 div from 10 MHz to
full bandwidth
< 5 mV/div: 1 div from DC to 10 MHz, 1.5 div from 10 MHz to
20 MHz
Table 11
Equipment Required for Trigger Sensitivity Test
Description
Critical Specifications
Recommended
Model/Part Numbers
Signal Generator
100 kHz to 1 GHz at 200 mVrms, 0.01 Hz
frequency resolution, jitter: < 2 ps
Keysight 8648A,
Keysight E4400B, or
Keysight N5181A.
BNC Cable
50
Ω
characteristic impedance, BNC (m)
connectors
Keysight 8120-1840,
Keysight 10503A
Type-N to BNC Adapter Type-N (m) to BNC (f)
Keysight 1250-0780
50
Ω
BNC Feed
Through Adapter
50
Ω
BNC (f) to BNC (m) feed through
terminator
Keysight 0960-0301
Содержание 1000B Series
Страница 1: ...Keysight 1000B Series Oscilloscopes Service Guide ...
Страница 6: ...6 Keysight 1000B Series Oscilloscopes Service Guide Contents ...
Страница 8: ...8 Keysight 1000B Series Oscilloscopes Service Guide Figures ...
Страница 10: ...10 Keysight 1000B Series Oscilloscopes Service Guide Tables ...
Страница 40: ...40 Keysight 1000B Series Oscilloscopes Service Guide 1 Service ...
Страница 44: ...44 Keysight 1000B Series Oscilloscopes Service Guide Index ...
Страница 45: ......