Operating Instructions
Operator’s check
The operator's check allows the operator to make a quick check of the solid state
switches prior to use or if a failure is suspected.
Description
The solid state switch is connected to a network analyzer configured for the
S-parameter measurement. The network analyzer may be set to sweep over the
whole or selected frequency range of the solid state switch to be verified. The S
21
(insertion loss) measurement is the best way to determine if the switch is faulty by
applying the appropriate logic to the CTRL pin.
Figure 3-1
Connection to perform quick check
CAUTION
The RF port center conductors connect directly to the GaAs MMIC and are
thus sensitive to electrostatic discharge (ESD). ESD exceeding the level
on page 3 or RF power applied is
greater than the maximum specified in
may cause permanent
damage to the device.
Network analyzer
Network analyzer
Solid state switch
Solid state switch
Keysight
U9397A/C
Keysight
U9397A/C
Содержание U9397A
Страница 1: ...Keysight U9397A C FET Solid State Switch SPDT Operating and Service Manual...
Страница 6: ...6 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 8: ...8 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 10: ...10 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...3 Operating Guide 26 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...