![Keysight Technologies U9397A Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/keysight-technologies/u9397a/u9397a_operating-and-service-manual_1962273003.webp)
Keysight U9397A/C Operating and Service Manual
3
Environmental Conditions
The table below shows the general environmental requirements for this
instrument.
Environmental condition
Requirement
Temperature
Operating condition:
– –40 °C to +65 °C
Storage condition:
– –65 °C to +85 °C
Cycling condition:
– –65 °C to +150 °C, 10 cycles at 20 °C per minute, 20 minutes dwell time
per MIL-STD-883F, Method 1010.8, Condition C (modified)
Humidity
Operating condition:
– 50% to 95% RH at 40 °C, one 24 hour cycle, repeated five times
Storage condition:
– <95% RH at 40 °C, five days
Shock
Half sine, smoothed:
– 1000 G at 0.5 ms, 3 shock pulses per orientation, 18 total per
MIL-STD-883F, Method 2002.4, Condition B (modified)
Vibration
Broadband:
– 50 to 2000 Hz, 7.0 G rms, 15 minutes, per MIL-STD-883F, Method
2026-1 (modified)
Altitude
Storage condition:
– <15300 meters (50000 feet)
ESD immunity
Direct discharge:
– 4 kV (to outer conductor) per IEC 61000-4-2
Air discharge:
– 8 kV (to center conductor) per IEC 61000-4-2
Содержание U9397A
Страница 1: ...Keysight U9397A C FET Solid State Switch SPDT Operating and Service Manual...
Страница 6: ...6 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 8: ...8 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 10: ...10 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 26: ...3 Operating Guide 26 Keysight U9397A C Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...