Performance Verification Test
Performance Verification Test
6
U5855A Service Guide
1
Set up the U5855A and the black body source as shown in
2
Ensure that the optical axis of the black body source is horizontal with
the U5855A.
3
Set the following parameters of the U5855A:
•
Emissivity (
ε
):
1
•
Object distance:
1 m
•
Ambient temperature:
25 °C
•
Humidity:
50%
•
Center spot:
ON
4
Ensure that the focus of the U5855A is targeted at the surface of the
black body source.
5
Ensure that the image of the black body source is displayed at the
center of the U5855A LCD screen.
shows an example of the
image when performing the verification.
Figure 1-2
Black body source image