3
Troubleshooting and Repair
Operating and Troubleshooting Hints
18
U5855A Service Guide
Operating and Troubleshooting Hints
If the U5855A fails to operate, check and replace the battery if necessary.
And if the U5855A still does not function properly, check the operating
procedures in the User's Guide.
The following table helps you identify some basic problems and their
solutions. Proceed with the described advice.
WA R N I N G
To avoid electrical shock, do not perform any servicing on the U5855A unless you are
qualified to do so.
Table 3-1
Troubleshooting hints
Malfunction
Probable cause
Troubleshooting procedure
The U5855A does not turn on.
Power is not supplied to the U5855A.
Check if the power cord is properly
connected to the AC power adapter.
Check if the power cord is properly
connected to the mains power outlet.
Check if the AC power adapter is operating
normally. If not, replace the adapter.
If the U5855A is running on battery, check
the battery condition.
Check if the battery is properly installed in
the U5855A.
If the battery is empty, charge the battery.
No thermal images displayed.
The external shutter is not opened
properly.
Open the external shutter fully and lock it
in place.
The infrared (IR) sensor fails to operate
properly.
Return the U5855A to Keysight for
servicing.
The thermal image is abnormal
(the previous image still remains).
The internal shutter is not opened
properly.
Check if there is a sound of the internal
shutter closing when switching from the
freeze mode to the run mode. If no, return
the U5855A to Keysight for servicing.
Содержание U5855A
Страница 1: ...Keysight U5855A TrueIR Thermal Imager Service Guide...
Страница 2: ......
Страница 3: ...U5855A Service Guide I...
Страница 14: ...XII U5855A Service Guide Obtaining Repair Service Worldwide 26...
Страница 16: ...XIV U5855A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 18: ...XVI U5855A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...