38 Keysight U3045AM04 User’s and Service Guide
U3045AM04
Address and Data Values
1-
Address and Data Values
Setting the Test Port Paths with Address and Data
Refer to
for information to set the internal
switch paths of the test set.
The address is the first value in the Test Set I/O control or GPIB data command.
The second value controls the source and receiver paths of the ports. To select
a test set port configuration both switches must be set to complete the source
and receiver paths. To do this you must add the data values together to
determine the data command value for each port, which is the second number
in the Test Set I/O or GPIB command. Refer to
and in
“GPIB Control Mode” on page 35
.
Example 1: Refer to
If the ports have different addresses, two separate address data commands
must be used.
Port 5 is the Source and Port 6 is the Receiver.
Source Port 5 = address 0, data 1 and Receiver Port 6 = address 16, data 16.
Two separate commands must to be sent, you may use the same dialog box.
The entry will be 0.1 in one command line, and 16.16 in the second line.
Example 2:
If the ports have the same address, only one command is needed.
Port 1 is the Source and Ports 5 is the Receiver.
Source Port 1 = address 0, data 0 and Receiver Port 5 = address 0, data 16. The
data values are added together, the entry will be 0.16.
Table 8
Test Port Address and Data Values
Address
Source Path
Receiver Path
Data
0
1
0
16
0
Ports
1
5
1
5
16
Ports
3
6
3
6
32
Ports
4
7
4
7
64
Ports
2
8
2
8