58
Keysight U3042AM20 User’s and Service Guide
U3042AM20
Control Lines
-
Setting the Control Lines with Address and Data Values
This section describes how to control the DUT control lines. Refer to
Setting the PNA to Standalone Mode
1. Select Utility > System > Configure > Multiport Capability -OR-
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...)
.
In the dialog box, select Restart as a Standalone PNA > OK.
As described in
“Setting the Test Port Paths with Address and Data” on page 47
<address>.<data> values are determined in a similar manor, with the following
exceptions:
—
— The <address> value is always = 112
All DUT control lines remain in a high logic state until set by the address data 0 = logic
high. All control lines are set by the sum of the data sent to the Test Set.
Example: To set control line 1 and 8 to a low state send command 112.129 (1 + 128).
All other lines remain in a high state.
Table 15
Test Set DUT Control Address and Data Logic Table
Address
Data
Description
Pin
112
0
ALL the Control Lines set to logic high
13
112
255
ALL the Control Lines set to logic low
14
112
1
DUT Control Line 1 set to logic low
1
112
2
DUT Control Line 2 set to logic low
2
112
4
DUT Control Line 3 set to logic low
3
112
8
DUT Control Line 4 set to logic low
4
112
16
DUT Control Line 5 set to logic lowe
5
112
32
DUT Control Line 6 set to logic low
6
112
64
DUT Control Line 7 set to logic low
7
112
128
DUT Control Line 8 set to logic low
8