U3042AM08
Control Lines
Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
NOTE
The following control feature will function only while the analyzer is in Standalone
Mode.
Setting the Analyzer to Standalone Mode
1.
Select Utility > System > Configure > Multiport Capability
-OR-
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...)
.
In the dialog box
, select
Restart as a Standalone PNA > OK
As described in
“Address and Data Values” on page 5
, the <address>.<data> values
are determined in a similar man
ne
r, with the following exceptions:
•
• The <address> value is always = 112
Table 1
6
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
Description
Line/Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic high or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8
62
Keysight U3042AM08 User's and Service Guide
Содержание U3042AM08
Страница 7: ...U3042AM08...