U3022BH10
Safety and Regulatory Information
Before Applying Power
Verify that the premises electrical supply is within the range of the instrument. The
instrument has an autoranging power supply.
If this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a
normal condition (in which all means for protection are intact) only.
The Mains wiring and connectors shall be compatible with the
connector used in the premise electrical system. Failure, to ensure
adequate earth grounding by not using the correct components
may cause product damage, and serious injury.
This product is designed for use in Installation Category II and
Pollution Degree 2.
Verify that the premise electrical voltage supply is within the range
specified on the instrument.
Ventilation Requirements:
When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The
ambient temperature (outside the cabinet) must be less than the maximum
operating temperature of the instrument by 4 °C for every 100 watts
dissipated in the cabinet. If the total power dissipated in the cabinet is
greater than 800 watts, forced convection must be used.
118
U3022BH10 User's and Service Guide