U3022AM06
Installation Procedures, Test Set to Analyzer
Figure 25
illustrates the final two digits of the part number for each cable. The cables
must be connected in the numeric order listed in
.
8-Port Interface Connections
Table 10
8-Port RF Cable Connection (U3021-60043)
Numeric
Order
RF Cables
From: PNA_X
To: Test Set
1
U3022-20116
Port 1 SOURCE OUT
Ports 3, 5 and 7 SOURCE IN
2
U3022-20117
Port 1 CPLR THRU
Port 3, 5 and 7 CPLR THRU
3
U3022-20118
Port 1 CPLR ARM
Port 3, 5 and 7 CPLR ARM
4
U3022-20119
Port 1 RCVR A IN
Port 3, 5 and 7 RCVR OUT
5
U3022-20122
Port 2 CPLR ARM
Ports 4, 6 and 8 CPLR ARM
6
U3022-20120
Port 2 SOURCE OUT Port 3, 5 and 7 SOURCE IN
7
U3022-20121
Port 2 CPLR THRU
Port 3, 5 and 7 CPLR THRU
8
U3022-20123
Port 2 RCVR B IN
Port 3, 5 and 7 RCVR OUT
16
16
18
18 19 17
21
23 22
19 17
23 21
20
22
20
34
Keysight U3022AM06 User's and Service Guide