U3022AM06
Installation Procedures, Test Set to Analyzer
RF Interface Cable Connections (U3021-60043)
2-Port Analyzer
illustrates the cable configuration of the Test Set to the analyzer.
The cables have been supplied with Cable Kit (U3021-60043).
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x4)
on the PNA-L. The RCVR R1 to SOURCE OUT reference loop jumper (x1) remain on the
front panel.
2.
Connect the RF interconnect cables from the PNA to the Test Set in the order
listed. As you are connecting each cable, torque to 8 in-lb. Refer to
and
Over torque will cause damage to the Test Set and may cause the connectors to spin
or become loose.
Each end of the interconnect RF cables have a different length from the bend. When
connecting the RF Interconnect cables be sure that the longer end (from the bend) is
connected to the Analyzer.
Cable Connection Procedure
Keysight U3022AM06 User's and Service Guide
33