U3022AM06
Installation Procedures, Test Set to Analyzer
RF Interface Cable Connections (U3021-60045)
4-Port Analyzer
illustrates the cable configuration of the Test Set to a 4-
Port analyzer. The cables have been supplied with kit (U3021-60045).
Cable Connection Procedure
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x4)
on the PNA-L. The RCVR R1 to SOURCE OUT reference loop jumper (x1) remain on the
front panel.
2.
Connect the RF interconnect cables from the PNA-L to the Test Set. As you are
connecting each cable, torque to 8 in-lb. Refer to
and
Over torque will cause damage to the Test Set and may cause the connectors to spin
or become loose.
Each end of the interconnect RF cables have a different length from the bend. When
connecting the RF Interconnect cables be sure that the longer end (from the bend) is
connected to the Analyzer.
Table 9
10-Port Interface Cable Connection (U3021-60045)
RF Cables
From: PNA-L
To: Test Set
Z5623-20418
SOURCE OUT
SOURCE OUT
Z5623-20418
CPLR ARM
CPLR ARM
Z5623-20419
CPLR THRU
CPLR THRU
Z5623-20419
RCVR IN
RCVR OUT
Keysight U3022AM06 User's and Service Guide
31