Capacitive Load Boundary for Voltage Priority Mode
The following figures show the boundary limitations for capacitive loads with equivalent series
resistance. Operation
below
the two marginal stability lines when operating with either the High or
Low
may result in output instability.
500 V units
950 V units
500 V, 10 kW unit
950 V, 10 kW unit
500 V, 5 kW unit
Voltage Programming Response (All Models)
The following graph shows the output voltage programming response characteristic. This applies for
small signals only, under no load conditions.
1 Quick Reference
58
Keysight RP7900 Series Operating and Service Guide