Verification Setups
Test Considerations
For optimum performance, all verification and calibration procedures should comply with the
following:
l
Ambient temperature is stable, between 18 and 28 °C.
l
Ambient relative humidity is less than 80%.
l
30 minute warm-up period before verification or adjustment.
l
Cables as short as possible, twisted or shielded to reduce noise.
6 Verification and Calibration