SCPI Error Messages
The Keysight instrument returns error messages in accord with the SCPI standard.
l
Up to 20 errors can be stored in each interface-specific error queue (one each for GPIB, USB, VXI-
11, and Telnet/Sockets.) Errors appear in the error queue of the I/O session that caused the error.
l
The front-panel ERR annunciator turns on when there are one or more errors are in the error
queue.
l
A special global error queue holds all power-on and hardware-related errors (for example, over-
temperature).
l
Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read them. Once you have
read all interface-specific errors, the errors in the global error queue are retrieved. When you have
read all errors from the error queue, the ERR annunciator turns off.
l
If more than 20 errors have occurred, the last error stored in the queue (the most recent error) is
replaced with -350,"Error queue overflow". No additional errors are stored until you remove errors
from the queue. If no errors have occurred when you read the error queue, the instrument
responds with +0,"No error".
l
The front panel reports errors from all I/O sessions and the global error queue. To read the error
queue from the front panel, press the ERROR key.
l
Error conditions are also summarized in the Status Byte Register. See
for details
l
The interface-specific error queues are cleared by power cycles and *CLS. The error queue is not
cleared by *RST.
l
SCPI:
SYSTem:ERRor?
Read and clear one error from the queue
Errors have the following format (the error string may contain up to 255 characters):
Error Device-dependent Errors (these errors set Standard Event Status register bit #3)
0 No error
This is the response to the ERR? query when there are no errors.
101 Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.
102 Calibration password is incorrect
The calibration password is incorrect.
103 Calibration is inhibited by switch setting
5 SCPI Programming Reference
252
Keysight RP7900 Series Operating and Service Guide
Содержание RP7900 Series
Страница 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Страница 2: ......
Страница 51: ...Dimensions RP793xA and RP794xA Keysight RP7900 Series Operating and Service Guide 51 1 Quick Reference ...
Страница 60: ...Dimensions RP795xA and RP796xA 1 Quick Reference 60 Keysight RP7900 Series Operating and Service Guide ...
Страница 100: ......
Страница 244: ...Status Diagram 5 SCPI Programming Reference 244 Keysight RP7900 Series Operating and Service Guide ...
Страница 246: ...Trigger Diagram 5 SCPI Programming Reference 246 Keysight RP7900 Series Operating and Service Guide ...
Страница 262: ......
Страница 314: ......