Find us at www.keysight.com
Page 13
Calibration kits
Coaxial measurements
Mechanical calibration kits include standards, such as opens, shorts and loads, which are measured by
the network analyzer for increased measurement accuracy.
Choose a calibration kit for each connector type to be used.
Economy
, includes:
•
Open standards (male and female)
•
Short standards (male and female)
•
Fixed-termination standards (male and female)
Standard
, includes the devices in the economy kit and adds:
•
Sliding load standards (male and female) or a series of offset shorts
Precision
, includes the devices in the economy kit and adds:
•
50 Ω airline(s) for TRL calibration
•
TRL adapters
Electronic calibration (ECal) kits replace mechanical calibration standards with one solid-state calibration
module that is controlled by the network analyzer via USB, to present many different impedances to the
test ports. A full two-port calibration can be performed quickly with a single connection. This technique
reduces operator errors and connector wear and abrasion.
For more information about ECal modules, refer to the technical overview 5963-3743E (N4690, 85090,
N4430 and N7550 Series).
Waveguide measurements
For waveguide measurements, Keysight offers mechanical calibration kits that include:
•
Waveguide-to-coax adapters (X, P, K, R, Q, U, V)
•
Precision waveguide section
•
Flush short circuit
•
Fixed terminations
•
Straight section