Installation Procedure
12.
Refer to
. Position the new A13A1
(1)
on the spacer such that the holes in the A13A1
housing align with the holes in the spacer. Screws in only three of the four holes in the A13A1
housing will be used to secure A13A1 to the spacer. No screw will be used in the hole in the
lower right corner of A13A1. Secure A13A1
(1)
to the spacer using three 0515-0661 screws
(2)
and torque to 3 inch-pounds using the T-6 driver.
Figure 14
Mount New A13A1 Bandpass Filter
13.
Connect A13W1 to A13J10 and torque to 10 inch-pounds.
14.
Connect A13W2 to A13J11 and torque to 10 inch-pounds.
15.
Torque the SMA connectors on A13A1 to 10 inch-pounds.
16.
Reconnect ribbon cable W17 to A13J12.