List Commands
List commands program an output sequence of multiple voltage or current settings. A comma-
delimited list of up to 512 steps may be programmed. Note that these commands only apply in the
presently active priority mode, either voltage priority or current priority.
[SOURce:]LIST:COUNt <
value
>, (@<chanlist>)
[SOURce:]LIST:COUNt? [MIN|MAX,] (@<chanlist>)
Sets the list repeat count. This sets the number of times that a list is executed before it completes. The
count range is 1 through 4096. Use the Infinity parameter to run the list continuously.
Parameter
Typical Return
1 – 4096|MIN|MAX|INFinity, *RST 1
<list count>
Sets the list count to 10: LIST:COUN 10, (@1)
l
Use
to stop the list at any time. When the list is aborted, the output returns to
the settings that were in effect before the list started.
l
The query returns the list count for each channel specified. If a repeat count of 9.9E37 is returned,
it means the list is set to repeat continuously.
[SOURce:]LIST:CURRent[:LEVel] <
value
>{,<
value
>}, (@<chanlist>)
[SOURce:]LIST:CURRent[:LEVel]? (@<chanlist>)
[SOURce:]LIST:VOLTage[:LEVel] <
value
>,{<
value
>}, (@<chanlist>)
[SOURce:]LIST:VOLTage[:LEVel]? (@<chanlist>)
Specifies the setting for each list step. Values are specified in either amperes or volts. A comma-
delimited list of up to 512 steps may be programmed.
Parameter
Typical Return
0 to maximum. The maximum value depends on rating of the power mod-
ule.
*RST 1 step set to the minimum programmable value.
<list value 1>,
<list value 2>,
<list value 3>...
Programs a current list of 3 steps. LIST:CURR 3,2,1, (@1)
Programs a voltage list of 3 steps. LIST:VOLT 20,10,5, (@1)
l
The order in which the list values are entered determines the sequence when the list executes.
l
To create a valid list, the Voltage, Current, BOST, EOST, and Dwell lists must all be the same
length or have a length of 1, which is interpreted as having the same length as the longest list.
l
This command overwrites any previously programmed list; it does not append to the previous list.
7 SCPI Programming Reference
296
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......