[SOURce:]ARB:CURRent:EXPonential:TCONstant <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:EXPonential:TCONstant? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:EXPonential:TCONstant <
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:EXPonential:TCONstant? [MIN|MAX,] (@<chanlist>)
Specifies the time constant of the exponential curve. Referenced to tc in the Exponential diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:TCONstant <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:TCONstant ? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:EXPonential:TCONstant <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:EXPonential:TCONstant ? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the time constant of the exponential curve for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 1
<time constant>
Step 0 - 99, *RST 0
(none)
Programs a time constant for the current Arb: ARB:CURR:EXP:TCON 5,(@1)
Programs a time constant for the voltage Arb: ARB:VOLT:EXP:TCON 5,(@1)
Programs a time constant for step 0: ARB:SEQ:STEP:CURR:EXP:TCON 5, 0,(@1)
[SOURce:]ARB:CURRent:EXPonential:TIMe <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:EXPonential:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:EXPonential:TIMe <
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:EXPonential:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the total time for the current or voltage to go from the starting level to the ending level in
seconds. Referenced to t1 in the Exponential diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:TIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:EXPonential:TIMe<
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:EXPonential:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the total time for the current or voltage to go from the starting level to the ending level in
seconds for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 0
<total time>
Step 0 - 99, *RST 0
(none)
Programs the total time for the current Arb: ARB:CURR:EXP:TIM 10,(@1)
Programs the total time for the voltage Arb: ARB:VOLT:EXP:TIM 10,(@1)
Programs the total time for step 0: ARB:SEQ:STEP:CURR:EXP:TIM 10, 0,(@1)
7 SCPI Programming Reference
240
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......