Keysight N5990A MIPI M-PHY Receiver Test User Guide
4
Calibrations
Level Calibration Terminated / Into Open
This calibration is available for all hardware configurations.
Figure 43
Level Calibration Terminated
Purpose
This procedure is used to calibrate the amplitude and offset of the signal
generators when those are terminated or into open.
This calibration should run once at each selected lane (see
For the M8020A/ DSGA
setup, the calibration for Data0 includes
the switch. If more than one channel is selected, the other data outputs
are calibrated without switch.