44
Keysight N5990A MIPI M-PHY Receiver Test User Guide
2
N5990A M-PHY Station
Applied Data Rate, Multiplier Factor, Applied Frequency
These properties show the setting that will be applied during testing.
If
Nominal Data Rate
is selected, the
Applied Data Rate
is equal to the
Custom Data Rate
, but the
Applied Frequency
can be sightly different
from the
Ref.Clock Frequency
selected.
If
Nominal Ref. Clock
is selected, the
Applied Frequency
is equal to the
Ref. Clock Frequency
selected, but the
Applied Data Rate
can be sightly
different from the
Custom Data Rate
.
Default Levels
Click
Default Levels
in the
Configure DUT
Panel to display the
Set M-PHY
Default Levels
window (see
). It allows the levels (
Amplitude
and
Offset
) for the
Data
,
DSGA
and
Reference Clock
channels to be set.
Figure 15
Set M-PHY Default Levels panel