Keysight N5990A MIPI M-PHY Receiver Test User Guide
109
Calibrations
4
Short Term RJ Calibration
It is available for all hardware configurations. It is only defined for
Gear1
and
Gear2.
Figure 63
Short Term RJ Calibration
Purpose
The procedure calibrates the high frequency RJ which is generated
internally by the BERT.
This calibration should run once at each HS-G1/G2 data rate (see
). For spec 4.1. this calibration is only mandatory for B-series (for
A-series the procedure is shown in the test tree as informative).