![Keysight Technologies N5224BU-601 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/keysight-technologies/n5224bu-601/n5224bu-601_installation-note_1961889033.webp)
Installation Note N5225-90117 33
Description of the Upgrade
Installation Procedure for the Upgrade
— W18 (reuse) (N5245-20111)A32 port 2 receiver coupler to W17
— W38 (N5245-20037)Port 3 RCVR C IN to A28 mixer brick (C)
— W44 (N5245-20020)REF 3 RCVR R3 IN to A28 mixer brick (R3)
— W12 (reuse) (N5245-20109)A29 port 1 receiver coupler to W11
— W20 (N5245-20099)Port 1 CPLR THRU to A33 port 1 coupler
— W24 (N5245-20098)Port 3 CPLR THRU to A34 port 3 coupler
— W26 (N5245-20015)A34 port 3 coupler to front-panel Port 3 CPLR ARM
— W23 (N5245-20051)A30 port 3 rcvr coupler to front-panel Port 3 SOURCE
OUT
— W25 (N5245-20016) A30 port 3 rcvr coupler to front-panel REF 3 SOURCE
OUT
, install clamp (part number 1400-1331) to secure W25
(N5245-20016).
Figure 18
Location of Cable Clamp for W25 (N5245-20016)
— W22 (N5245-20014)A33 port 1 coupler to front-panel Port 1 CPLR ARM
— W21 (N5245-20110)A29 port 1 receiver coupler to A37 reference mixer
switch
* If the screws that attach the reference mixer switch to the test set deck
were loosened, torque these screws now to 9 in-lbs.
* As shown in
, install two clamps (part number 1400-1334) to
secure W21 (N5245-20110).
Содержание N5224BU-601
Страница 43: ...Installation Note Xxxxx xxxxx 3...