Keysight N5224A/25A Service Guide 4-41
Troubleshooting
Measurement System Troubleshooting
Checking Port 1, 2, 3, or 4 Power Outputs (A, B, C, or D Signals)
The object of this check is to verify the power of the output signal across the
entire frequency range. Perform this test if there is an observed problem only
with one receiver trace. The ten second sweep is slow enough to allow you to
observe the output power on the power meter as the sweep occurs.
1.
Connect the power sensor to the suspect port.
2.
Set the trace to measure S
11
, S
22
, S
33
, or S
44
, dependent on the suspect
port.
3.
Observe the power reading on the power meter as the sweep occurs on
the analyzer.
4.
The measured output power on the power meter should be at least the
preset power level ±1 dB over the entire frequency range.
—
If the measured power is correct, go to
.
—
If the measured power is not correct, go to
through the Signal Separation Path” on page 4-41
.
Checking the Signal through the Signal Separation Path
For all of the following checks, refer to the block diagrams at the end of this
chapter and to any of the following that are appropriate:
—
“2-Port Configuration, Option 200” on page 6-22
—
“Bottom RF Cables, Standard 2-Port Configuration, Option 200” on
—
“2-Port Configuration, Option 219” on page 6-47
—
“Bottom RF Cables, 2-Port, Option 219” on page 6-49
—
“4-Port Configuration, Option 400” on page 6-54
—
“Bottom RF Cables, Standard 4-Port Configuration, Option 400” on
—
“4-Port Configuration, Option 419” on page 6-84
—
“Bottom RF Cables, 4-Port, Option 419 (Ports 1 and 2)” on page 6-86
—
“Bottom RF Cables, 4-Port, Option 419 (Ports 3 and 4)” on page 6-89
Trace loss in the signal separation group is due to one or more of the following
assemblies being defective:
— A29, A30, A31, or A32 receiver coupler
— A38, A39, A40, or A41 source step attenuator
— A42, A43, A44, or A45 bias tee
— A33, A34, A35, and A36 test port coupler
Содержание N5224A
Страница 124: ...4 52 Keysight N5224A 25A Service Guide Troubleshooting Instrument Block Diagrams Sheet 2 4 ...
Страница 132: ...5 8 Keysight N5224A 25A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 149: ...Keysight N5224A 25A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 6 4 Port Receiver Group ...
Страница 150: ...5 26 Keysight N5224A 25A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 7 2 Port Receiver Group ...
Страница 266: ...6 110 Keysight N5224A 25A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 346: ...8 18 Keysight N5224A 25A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...