Keysight N5290/1A Service Guide 3-5
System Tests and Adjustments
Before You Begin
Protect Against Electrostatic Discharge (ESD)
This is important. If not properly protected against, electrostatic discharge can
seriously damage your analyzer, resulting in costly repair.
Review the Principles of Connector Care
Proper connector care and connection techniques are critical for accurate and
repeatable measurements. Refer to
for tips on connector care.
Prior to making connections to your analyzer, carefully review the information
about inspecting, cleaning, and gaging connectors. Refer to the calibration kit
documentation for detailed connector care information.
For course numbers about additional connector care instruction, contact
Keysight Technologies. Refer to
“Contacting Keysight” on page 2-7
To reduce the chance of electrostatic discharge, follow all of the
“Electrostatic Discharge Protection” on
, for all of the procedures in this chapter.
To achieve the maximum system stability, allow the analyzer to warm up
for at least 90 minutes.
Cleaning connectors with alcohol shall only be done with the
N5xxxB removed from the system, and in a well-ventilated area.
Allow all residual alcohol moisture to evaporate and the fumes to
dissipate prior to reconnecting the N5xxxB to the system.
Table 3-1
Connector Care Quick Reference Guide
Handling and Storage
Do
— Keep connectors clean
Do Not
— Touch mating-plane surfaces
— Extend sleeve or connector nut
— Set connectors contact-end down
— Use plastic end-caps during storage
— Store connectors or adapters loose
Visual Inspection
Do
— Inspect all connectors carefully
Do Not
— Use a damaged connector - ever
— Look for metal particles, scratches, and
dents
Connector Cleaning
Do
— Try compressed air first
Do Not
— Use any abrasives
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...