6
Firmware Upgrade Guide N5191-90023
Firmware Upgrade Guide
Step 2. Run the Signal Generator’s Functionality Check
Use the following procedure to confirm that the signal generator powers up and that the internal
check identifies no errors. The internal check evaluates the operation of the signal generator and
returns an error message if it detects a problem.
1.
Turn on power to the signal generator and let it warm up for at least five minutes.
2.
Check to see if the
ERR
annunciator is on.
—
If the
ERR
annunciator is
off
, the signal generator’s functionality check passed.
—
If the
ERR
annunciator is
on
:
a.
View the error queue:
Press
Error
.
b.
Resolve all problems causing errors.
If you are unable to resolve the errors, contact Keysight Technologies (see
c.
Clear the error queue: Press
Clear Error Queue(s)
3.
Repeat the functionality check:
Cycle the signal generator’s power.
4.
Repeat
.