Firmware Upgrade Guide N5191-90023

Firmware Upgrade Guide

Step 2. Run the Signal Generator’s Functionality Check

Use the following procedure to confirm that the signal generator powers up and that the internal 

check identifies no errors. The internal check evaluates the operation of the signal generator and 

returns an error message if it detects a problem.

1.

Turn on power to the signal generator and let it warm up for at least five minutes.

2.

Check to see if the 

ERR

 annunciator is on.

If the 

ERR

 annunciator is 

off

, the signal generator’s functionality check passed.

If the 

ERR

 annunciator is 

on

:

a.

View the error queue: 
Press 

Error

b.

Resolve all problems causing errors.
If you are unable to resolve the errors, contact Keysight Technologies (see 

page 11

).

c.

Clear the error queue: Press 

Clear Error Queue(s)

 and go to 

step 3

.

3.

Repeat the functionality check: 
Cycle the signal generator’s power.

4.

Repeat 

step 2

.

Содержание N5191A

Страница 1: ...For the latest revision of this installation note go to the following website http www keysight com find uxg Keysight Technologies N5191A 93A Agile Signal Generators Firmware Upgrade Guide...

Страница 2: ...to the implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in c...

Страница 3: ...re Upgrade Guide Product Affected N5191A 93A Serial Numbers All Options All To Be Performed By X Keysight Technologies Service Center X Personnel Qualified by Keysight Technologies X Customer Estimate...

Страница 4: ...operation of the signal generator If you are not comfortable with using the signal generator s menus and with entering parameter values refer to the instrument s User s Guide and familiarize yourself...

Страница 5: ...4 01 or later Netscape Navigator 4 0 or later If using GPIB an installed and configured GPIB IO interface card and a GPIB cable either the Keysight or National Instruments I O Libraries If using LAN...

Страница 6: ...roblem 1 Turn on power to the signal generator and let it warm up for at least five minutes 2 Check to see if the ERR annunciator is on If the ERR annunciator is off the signal generator s functionali...

Страница 7: ...it is recommended that you record the current LAN settings so you can restore them back to their original settings after upgrading the firmware Information on the LAN softkeys is provided in key Help...

Страница 8: ...see page 9 4 After the upgrade completes press the signal generator s Local hardkey to return front panel control 5 Run the signal generator s functionality check see page 6 If the upgrade is interru...

Страница 9: ...Firmware Upgrade Guide N5191 90023 9 Firmware Upgrade Guide Running the Upgrade Assistant...

Страница 10: ...10 Firmware Upgrade Guide N5191 90023 Firmware Upgrade Guide...

Страница 11: ...e regarding the signal generator s problem 2 Call the phone number listed on the Internet http www keysight com find assist that is specific to your geographic location If you do not have access to th...

Страница 12: ...This information is subject to change without notice Keysight Technologies 2018 Edition 1 July 2018 N5191 90023 www keysight com...

Отзывы: