Making Measurements
74
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
Starting the Lightwave Component Analyzer
1 Turn on the network analyzer and the optical test set.
2 Allow the Lightwave Component Analyzer to warm up. Both
the optical test set and the network analyzer need to stabilize
their internal temperature.
Please refer to “Measurement conditions” on page 132 for
information on settling.
3 If it does not start automatically, click on the icon on the
desktop to start the network analyzer application.
4. If it is not already running, start the LCA server by pressing
, or by clicking the LCA server icon on the desktop.
See “The LCA Function Keys” on page 58 for information on
displaying the LCA function keys.
N O TE
When starting the “LCA Measurement Setup” with no server running, the
LCA server is started automatically.