17
Displayed Average Noise Level (DANL)
Procedure
12.
Change the analyzer center frequency to the next value listed in
Press:
FREQ Channel
,
Center Freq
, [Table 2-1 Value],
GHz
13.
through
to fill in the remainder of
analyzer frequency range.
Table 2-1
Displayed Average Noise Level (DANL) Results
Center Frequency
Measured Average Noise Level (dBm)
Normalized Average Noise Level/(1 Hz
BW) (dBm)
Test Limits
(dBm)
10 MHz
–150
2 GHz
–150
6 GHz
–148
13 GHz
–147
20 GHz
–142
26.5 GHz
–135