Configuration and
Measurement Instructions

Keysight M9203A High-Speed
Digitizers Hardware Extension of
89600 VSA Software

Содержание M9203A

Страница 1: ...Configuration and Measurement Instructions Keysight M9203A High Speed Digitizers Hardware Extension of 89600 VSA Software...

Страница 2: ...and 27 405 3 and Department of Defense FAR Supplement DFARS 227 7202 the U S government acquires commercial computer software under the same terms by which the software is customarily provided to the...

Страница 3: ...Option Maximum Span Analysis Bandwidth M9203A SR2 625 MHz 1 25 GHz in interleaved mode The present document aims to describe how to configure and use VSA software with Keysight High Speed Digitizers...

Страница 4: ...9203A product page www keysight com find M9203A Technical Support Drivers Firmware and Software Once downloaded execute MD2DigitizersHwExt_Setup versions exe and follow the installation process VSA 89...

Страница 5: ...n the MD2 Digitizer Hardware Extension has to be re installed There are two solutions Option 1 From Windows Control Panel Programs Programs and Features select and right click on Keysight MD2 Digitize...

Страница 6: ...the system In order to proceed execute C Program Files Keysight MD2 89600 VSA Connectivity MD2Digitizers 89600VSAIntegration exe or using your installation path Follow the installation process Click...

Страница 7: ...tizer as the VSA measurement input by creating an Analyzer Configuration that uses your digitizer as the Logical Instrument ADC by per forming the following steps a From the menu go to Utilities Hardw...

Страница 8: ...n list 4 e You may either use the default name or specify another name for this ana lyzer configuration 5 f Click OK to close the New Hardware Configuration dialog 6 g Before closing the Hardware Conf...

Страница 9: ...to the User manual of your digitizer section Multi Purpose Inputs and Outputs Enable external clock See External Clock page 12 Enable interleaving See Interleaved Mode page 11 Error on overrange This...

Страница 10: ...e acquisition or digitizer parameter To proceed you have to enable the option MD2 Auto Calibration from Input Extensions window then to manually start the calibration once The following calibrations w...

Страница 11: ...has been ordered with the INT option Interleaving allows to double the sampling rate with the trade off of halving the available center frequency range as shown in the table below Model Option Mode C...

Страница 12: ...ent as shown in the table below Models Option Frequency Range M9203A SR2 1 8 to 3 2 GHz To use this option first connect an external clock signal in the accepted range to the CLK IN connector on the d...

Страница 13: ...nt Calibration Whenever the digitizer configuration is changed an internal calibration is required if this new configuration has not yet been calibrated in the current session A manual calibration can...

Страница 14: ...ingle Ended The frequency measurement parameters can be configured from the MeasSetup Frequency menu Frequency The default setup is shown below as example Model Option Band Center Span Start Stop SR2...

Страница 15: ...r Can be configured by user Main Time Length Standard VSA parameter Can be configured by user Maximum value depends on the memory option of your module Center Frequency Range The range depends on whic...

Страница 16: ...digitizer state that was set before being disconnected Hardware Set up an Analyzer Configuration with the digitizer as the Logical Instrument as described in 89600 VSA Software Configuration page 7 Ov...