Keysight M9195A/B PXIe Digital Stimulus/Response Startup Guide
23
Step 6: Execute Automatic Corrections
Automatic Corrections procedure
1
Turn on the M9195A/B system, and allow it to warm up for a minimum of 30
minutes.
2
Remove all cables from the front of the M9195A/B.
3
Start the Soft Front Panel (SFP) by clicking
Start > All Programs > Keysight > MDsr
> MDsr SFP
.
4
The SFP will ask to connect to an instrument.
In the Connect to Instrument dialog (see
on page 19), connect to the
M9195A/B module for which you want to execute Automatic Corrections.
5
Select the
Instrument
tab in the Soft Front Panel as in
.
You must remove all cables from the M9195A/B before executing
Automatic Corrections. Failure to remove all cables will result in
inaccurate adjustment and could result in damage to the
M9195A/B and anything attached to the cables (such as your
DUT).
Figure 9
SFP Instrument tab