5
Characteristics
5.1
Performance Specification
The performance specification can be found in the Data Sheet of the M8195A at:
http://www.keysight.com/find/M8195A
5.2
Operating Environment
Storage Temperature
−40 °C to +70 °C
Operating Temperature
0 °C to 40 °C
Operating Humidity
5% to 80% relative humidity, non-condensing
Operating Altitude
Up to 2000 m
Installation
Category II
Pollution
Degree 2
5.3
General
Power consumption
180 W (nom), 65 GSa/s operation
Safety designed to
IEC61010-1, UL61010, CSA22.2 61010.1 tested
EMC tested to
IEC61326-1
Warm-up time
30 min
Calibration interval
2 years recommended
Cooling
Requirements
When operating the M8195A choose a location that provides at least 80 mm
of clearance at rear, and at least 30 mm of clearance at each side for the
AXIe chassis.