Verification
3
Keysight J7211A/B/C Operating and Service Manual
23
Instrument Settings
In order to perform quick test verification, the operator needs to ensure that instrument’s settings
are following recommendation settings as stated below.
Table 3-2
Instrument settings
Product model
J7211A
J7211B
J7211C
Start frequency
200 MHz
200 MHz
250 MHz
Stop frequency
6 GHz
18 GHz
26.5 GHz
Number of points
30
60
106
CAUTION
Do not connect measuring terminals to mains.