Replacing 6000L Assemblies
5
6000L Series Oscilloscopes Service Guide
85
To remove the power shaft
1
Perform the previous procedures:
•
•
•
2
Use a flat-blade screwdriver to gently spread the latch while pushing the power
shaft forward.
Figure 22
Removing the power shaft.
3
Lift and remove it from the deck.
4
When reinserting the power shaft, push the shaft into the power switch until the
shaft snaps onto the switch.
CAUTION
Twisting the latch too much could cause it to break!