76
6000L Series Oscilloscopes Service Guide
5
Replacing 6000L Assemblies
To remove the system board
1
Perform the previous procedures:
•
2
Remove the intensity knob as described in step 2 on
.
3
Disconnect the front panel ribbon cable from the system board by pressing the
release tabs.
4
Disconnect the fan cable.
5
Remove the 3 hex nuts and washers from the rear BNC connectors using a
5/8-inch socket driver.
Figure 16
Preparing to remove the system board
T10 Screws
Fan Cable
System Board
Backlight Inverter Cable
Power Supply
Cable
Front Panel
Cable
Power
Supply
Cable
Hex
Standoffs