:SBUS<n> Commands
:SBUS<n>:LIN:TRIGger:PATTern:DATA
Command Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA <string>
<string> ::= "n" where n ::= 32-bit integer in unsigned decimal when
<base> = DECimal
<string> ::= "nn...n" where n ::= {0 | 1 | X | $} when
<base> = BINary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $} when
<base> = HEX
The :SBUS<n>:LIN:TRIGger:PATTern:DATA command specifies the LIN trigger data
pattern searched for in each LIN data field.
Set a <string> bit to "0" or "1" to set the corresponding bit in the data pattern to
low or high, respectively.
Set a <string> bit to "X" to ignore (mask off) that bit in the data pattern.
Use the "$" character to indicate that the value of the corresponding bit will not be
changed (the existing bit value is used).
When <base> = DECimal, the "X" and "$" characters cannot be entered. When
queried, the "$" character is returned when any bits in the pattern have the value
of "X" and <base> = DECimal. When any bits in a given nibble have the value of "X"
and <base> = HEX, the "$" character is returned for the corresponding nibble.
Query Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA?
The :SBUS<n>:LIN:TRIGger:PATTern:DATA? query returns the currently specified
LIN trigger data pattern.
Return Format
<string><NL>
NOTE
<base> is specified with the :SBUS<n>:LIN:TRIGger:PATTern:FORMat command. The default
<base> is BINary.
NOTE
The length of the trigger data value is determined by the
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth command.
NOTE
If more bits are sent for <string> than the specified trigger pattern data length, the most
significant bits will be truncated. If the data length size is changed after the <string> is
programmed, the added or deleted bits will be added to or deleted from the least significant
bits.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...