642
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
24
:POWer Commands
:POWer:ONOFf:THResholds
Command Syntax
:POWer:ONOFf:THResholds <type>,<input_thr>,<output_thr>
<type> ::= {ON | OFF}
<input_thr> ::= percent from 0-100 in NR1 format
<output_thr> ::= percent from 0-100 in NR1 format
The :POWer:ONOFf:THResholds command specifies the input and output
thresholds used in the Turn On/Turn Off analysis.
Turn On analysis determines how fast a turned on power supply takes to reach
some percent of its steady state output. Turn on time is the time between T2 and
T1 where:
• T1 = when the input voltage first rises to some percent (typically the 10%
threshold) of its maximum amplitude.
• T2 = when the output DC voltage rises to some percent (typically the 90%
threshold) of its maximum amplitude.
Turn Off analysis determines how fast a turned off power supply takes to reduce its
output voltage to some percent of maximum. Turn off time is the time between T2
and T1 where:
• T1 = when the input voltage last falls to some percent (typically the 10%
threshold) of its maximum amplitude.
• T2 = when the output DC voltage last falls to some percent (typically the 10%
threshold) of its maximum amplitude.
Query Syntax
:POWer:ONOFf:THResholds? <type>
The :POWer:ONOFf:THResholds? query returns the input and output threshold
settings for the turn on/turn off analysis type.
Return Format
<input_thr>,<output_thr><NL>
<input_thr> ::= percent from 0-100 in NR1 format
<output_thr> ::= percent from 0-100 in NR1 format
See Also
•
":POWer:SIGNals:VSTeady:ONOFf:OFF"
•
":POWer:SIGNals:VSTeady:ONOFf:ON"
•
•
•
•
•
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...