:FUNCtion<m> Commands
17
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
379
:FUNCtion<m>:DISPlay
Command Syntax
:FUNCtion<m>:DISPlay <display>
<m> ::= 1 to (# math functions) in NR1 format
<display> ::= {{1 | ON} | {0 | OFF}}
The :FUNCtion<m>:DISPlay command turns the display of the function on or off.
When ON is selected, the function operates as specified by the other
:FUNCtion<m> commands.
One math function can be displayed at a time. If one math function is on and then
another math function is turned on, the first math function will be turned off.
When math functions are off, they are still calculated so that they can be
cascaded, that is, used as a source for a higher math function.
Query Syntax
:FUNCtion<m>:DISPlay?
The :FUNCtion<m>:DISPlay? query returns whether the function display is on or
off.
Return Format
<display><NL>
<display> ::= {1 | 0}
See Also
•
"Introduction to :FUNCtion<m> Commands"
•
•
•
•
•
NOTE
Automatic vertical scaling of the math function waveform occurs when the function display is
turned from off to on. If you wait for the operation to complete (with an *OPC? query for
example), then query the math functions's scale, you can see the vertical scaling that was
automatically determined.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...