50
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Test External Trigger Sensitivity
This test applies to all models.
Verify the external trigger sensitivity at these settings:
•
100 MHz, 200 mV
pp
•
200 MHz, 350 mV
pp
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the power splitter input.
b
Connect one output of the power splitter to the Aux Trig input through a 50
Ω
feedthrough termination.
c
Connect the power sensor to the other output of the power splitter.