Troubleshooting
Front Panel LED Troubleshooting
If you see a failure with the on/off switch backlight LED, replace the On/Off board.
If the LED will still not illuminate, replace the motherboard and see if this fixes the
problem. If the LED still does not work, check the on/off cable connecting the
On/Off board to the motherboard, and check the connections of all cables on the
on/off board and the motherboard.
If any of the other LEDs on the front panel are not working, the first step is to
troubleshoot the acquisition assemblies to make sure they are working. If they are
working, replace the keyboard assembly and see if the LEDs will illuminate. If the
LEDs will not illuminate, the last step is to reload the oscilloscope software.
When reassembling the oscilloscope, be sure to reinstall all boards that were
replaced, but that were not causing the problem.