In the following section, we will describe the step by step process of using IQTools
to manually generate a FMCW waveform with the specifications described in
.
1
Connect the instruments as described in
The USB connections to the N5183B are not required since we will be using
controlling the MXG manually. If the system it being started up for the first
time, it is recommended to connect the IF and LO cables to the N9029ACST
upconverter only after the MXG and AWG has been set up correctly. This is to
ensure no accidental RF overload of the sensitive millimeter wave equipment
occurs.
2
Apply power to the instruments, upconverter and other accessories.
3
Set the RF frequency and Amplitude of the N5183B:
a
Press the
RF On/Off
button on the MXG to turn
off
the RF output first.
b
Press the
FREQ
button and enter
37.5 GHz
using the numeric keypad.
c
Press the
AMPTD
button, enter
6 dBm
using the numeric keypad.
Table 1
FMCW waveform parameters
Parameter Setting
Value
Comments
Radar waveform type
Linear FM, increasing.
Pulse period = 2 us.
Radar waveform frequency
77 to 81 GHz
IF frequency (M8195A)
2 to 6 GHz
N5183B frequency
37.5 GHz
Fc = 75 GHz because
N9029ACST operate on N = 2
harmonic factor. Minimum image
separation is 4 GHz and
adequately filtered by the
bandpass filter.
N5183B amplitude
4 dBm + loss of LO cable.
Содержание E8740A-070
Страница 1: ...Keysight E8740A 070 Performance Radar Signal Generation Solution Start up Guide...
Страница 4: ...4 Keysight E8740A 070 Start up Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 6: ...6 Keysight E8740A 070 Start up Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 15: ...Keysight E8740A 070 Start up Guide 15 6 You may need to reboot the system after installation...
Страница 50: ...50 Keysight E8740A 070 Start up Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...