Chapter 4
53
Preparation for Accurate Measurement (Executing Correction)
Executing OPEN/SHORT/LOAD Correction
4. Pr
ep
ar
at
ion
fo
r
Accu
ra
te
M
eas
ur
em
en
t
(Executing
C
orrection)
The load correction is performed at the range specified by
[:SENSe][:FIMPedance]:RANGe[:UPPer] command on page 262
, when the function
is set to OFF.
Defining LOAD correction standard
Before measuring LOAD correction data, you must define the LOAD correction standard.
To define the LOAD correction standard, use the following commands:
•
[:SENSe]:CORRection:CKIT:STAN3[:DATA] on page 231
•
[:SENSe]:CORRection:CKIT:STAN3:FORMat on page 232
Reading/writing correction data (saving/recalling correction
conditions)
After saving the correction conditions to a file, you can recall these conditions on the
E4981A at any time from the file.
The correction data is set up in the parameter-oriented format shown in the table below:
To read/write the correction data, use the following command.
•
[:SENSe]:CORRection:DATA on page 236
The correction data written using the above command is handled as follows:
• In case of STAN3: Saved as LOAD correction data for the measurement frequency
specified when the command is executed.
• In case of STAN1/STAN2: Saved as OPEN/SHORT correction data for all frequencies.
• Saved as the data for normal operation when the multi-correction function is OFF and
as the data for the multi-correction when it is ON (for the channel that has been selected
at execution).
Type of correction
Parameter format
OPEN correction
G-B or Cp-G
(select with the
[:SENSe]:CORRection:CKIT:STAN1:FORMat
command)
SHORT correction
R-X or Ls-Rs
(select with the
[:SENSe]:CORRection:CKIT:STAN2:FORMat
command)
LOAD correction
Parameter format to define the LOAD correction standard
(select with the
[:SENSe]:CORRection:CKIT:STAN3:FORMat
command)
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...